266 i Series / iX Series
Sub-Tab
Description
specifications. These limits must be entered manually into the
provided template file named “MS704_Distortion_Limits.ini”
located in the CIGui32 application directory. If the limits are
left at 0.00, no pass/fail criteria will be applied.
P/F
Pass or Fail indication.
A phase selection is available to the left of the graphs and table, which allows each
phase to be displayed, one at a time.
Additional user controls located on the EUT Performance Measurements Tab:
Control / Display
Description
Update Meas.
Triggers a new acquisition cycle and updates measurement data for all tabs. For
table SVF and TVF modes (Variable Frequency), data will be acquired at the
frequency set by the Test Frequency Dropdown control.
This button is always available regardless of what sub tab is selected.
Settings
Brings up a dialog box that allows scaling and color changes for the Waveform or
Current harmonics graphs. Note that setting changes will apply to graphs for all
phases.
For current harmonics graphs, only the absolute bar settings apply. The 704
measurement screen does not display relative current harmonic bar charts but
relative data is available in tabular format on the Cur Spectrum tab.
This button is available only if either the Waveforms or Cur Graph sub tab is
selected