User Manual
168 i Series / iX Series
Table A(NF)115:
Normal Operating Conditions (ac)
1 Phs Voltage and Frequency (ac)
100V to 122V, 360Hz to 650Hz
Apply for 30 mins each test
3 Phs Voltage and Frequency (ac)
100V to 122V, 360Hz to 650Hz
Apply for 30 mins each test
Emergency Operating Conditions (ac)
3 Phs Voltage and Frequency (ac)
100V to 122V, 360Hz to 650Hz
Apply for 30 mins each test
Normal Operating Conditions (ac)
Voltage Modulation 360Hz (ac)
Vamp = 0.18V to 1.24V, Mrate = 1Hz to 200Hz
Voltage Modulation 650Hz (ac)
Vamp = 0.18V to 1.24V, Mrate = 1Hz to 200Hz
Frequency Modulation 360Hz (ac)
Famp = 4.9 Hz to 0.2Hz, Mrate = 0.014Hz to 100Hz
Frequency Modulation 650Hz (ac)
Famp = 4.9 Hz to 0.2Hz, Mrate = 0.014Hz to 100Hz
Momentary Power Interruptions (ac)
Equipment with Digital Circuits
Momentary Power Interruptions (ac)
Equipment with Digital Circuits
Momentary Power Interruptions (ac)
Momentary Power Interruptions (ac)
Momentary Power Interruptions (ac)
Additional Requirement (Var Freq)
16.5.1.5.1 - Normal Voltage Surge (ac)
16.5.1.5.1 - Normal Voltage Surge (ac)
Normal Frequency Surge (ac)
16.5.1.7 - 1 Phs Voltage DC Content (ac)
Apply for 30 mins each test
16.5.1.7 - 1 Phs Voltage DC Content (ac)
Apply for 30 mins each test
16.5.1.7 - 3 Phs Voltage DC Content (ac)
Apply for 30 mins each test
16.5.1.7 - 3 Phs Voltage DC Content (ac)
Apply for 30 mins each test
8% +2%/-0% VTHD - Clipped Sine, 360Hz
Apply for 30 mins each test
8% +2%/-0% VTHD - Clipped Sine, 650Hz
Apply for 30 mins each test
Abnormal Operating Conditions (ac)
Apply for 5 mins each test
Apply for 5 mins each test
Apply for 5 mins each test
Apply for 5 mins each test
Momentary Undervoltage (ac)
Apply for 7 seconds each test